How to Read a HEPA/ULPA Factory Test Report
A useful HEPA or ULPA test report links one physical filter to one test result at stated conditions. Before checking the efficiency percentage, match the serial number, model, dimensions, airflow, test standard, and test date to the unit delivered. A high efficiency claim without that traceability is not enough for acceptance.
Factory testing and installed qualification answer different questions. ISO 29463-1:2024 covers classification, performance, testing, marking, and documentation of high-efficiency filters. ISO 29463-5:2022 specifies efficiency testing at the most penetrating particle size (MPPS). An installed system may still require leakage and airflow testing under ISO 14644-3 or a project-specific procedure.
Read the report in this order
Report field
What it tells you
What to verify
Filter ID and serial number
Which individual unit was tested
The number matches the filter label, packing list, and acceptance record.
Model and dimensions
The element’s construction and fit
Size, depth, frame, gasket or gel seal, and airflow direction match the approved drawing.
Standard and edition
The method used to classify and test the filter
The method matches the purchase specification; an old class label is not silently treated as an ISO 29463 class.
Rated test airflow
The flow at which results were measured
It matches the intended operating point or a justified test point.
Efficiency and MPPS
Particle removal at the stated test condition
The result and required class are both stated; ask how MPPS was determined.
Initial pressure drop
Resistance of a clean filter at the test airflow
Compare like-for-like flow and size, not isolated pressure numbers.
Local leak scan
Whether a localized defect was found
The scan method, pass criteria, result, and any permitted repair are documented.
Instrument and approval details
Traceability of the measurement
Date, instrument identification, calibration status, and authorized reviewer are present.
Do not confuse efficiency with a leak-free installation
Overall efficiency describes the proportion of test particles removed by the filter element under defined conditions. A local leak scan looks for a concentrated path through a defect or seal. A filter can have a convincing overall efficiency result yet still fail a local leak requirement. After installation, the frame, gasket, and housing become part of the system; the factory report does not certify those interfaces.
The term MPPS means the particle size that penetrates that filter medium most readily in the specified test. It is more useful than quoting an unqualified “0.3-micron efficiency” because test methods and filter media differ. Ask the supplier to identify the standard, the measured MPPS or method, and whether the stated result is overall or local efficiency.
Compare pressure drop fairly
Pressure drop depends on airflow. A lower number on one report is not necessarily a better filter if the two units were tested at different flows, face areas, or filter depths. When comparing suppliers, request pressure drop at the same specified airflow and confirm the dimensions and efficiency requirement are equivalent. The design team should also check that the fan has adequate margin at the operating condition.
For a replacement order, retain the old report and ask for a side-by-side comparison of filter size, seal type, rated flow, efficiency class, and clean pressure drop. This reveals substitutions that a headline class label can hide.
A five-minute acceptance check
Match the serial number on the report to the delivered unit.
Match the filter drawing, frame, and seal to the installation.
Confirm the required standard and class appear on the report.
Compare efficiency and pressure drop at the specified airflow.
Check the leak-scan result, test date, and instrument traceability.
If a field is absent, ask for the original test record or a corrected report. Do not fill the gap with a catalogue value from a different model.
Frequently asked questions
Is a certificate of conformity the same as an individual test report?
Not always. A certificate may state that a batch or product family conforms to a specification. For critical applications, request a serial-numbered report showing the measurements for the delivered element.
Does a factory report replace an on-site PAO test?
No. It verifies the element under factory test conditions. The installed test checks the assembled filter, seal, and housing at the site. See the installed HEPA validation overview.
Why might two H-class filters show different pressure drops?
Media area, pleat geometry, depth, and tested airflow may differ. Compare the same operating point and ask for the relevant performance curve, not just one catalogue number.
KLC’s HEPA/ULPA class comparison explains class terminology. For a purchase decision, request the actual model drawing and serial-numbered report alongside that general guide.
Technical sources
• ISO 29463-1:2024 — Classification, performance, testing and marking
• ISO 29463-5:2022 — Test method for filter elements at MPPS
• ISO 14644-3:2019 — Cleanroom performance test methods
La Exposición de Baterías y Tecnología de Vehículos Eléctricos e Híbridos 2025, un evento anual muy esperado para la industria global de nuevas energías, se celebró con éxito en Estados Unidos el 9 de octubre. Como empresa líder en filtración de aire y soluciones para salas blancasKLC participó en la exposición, presentando tecnología de vanguardia, soluciones profesionales y un profundo conocimiento del sector. Trabajamos con clientes y socios globales para demostrar con éxito nuestro valor clave en el apoyo a la cadena de suministro de fabricación de vehículos eléctricos y baterías. Filtros de aire de taller de batería de alta precisiónPresentamos filtros HEPA/ULPA de alta eficiencia para controlar el entorno de producción de baterías. Estos productos eliminan eficazmente el polvo fino y las partículas metálicas del aire, garantizando una producción de baterías extremadamente limpia y un rendimiento constante y seguro desde la fuente, lo que ha atraído la atención de los fabricantes de baterías. Intercambios profesionales, perspectivas de la industria El stand de KLC estuvo repleto de visitantes durante toda la exposición. Participamos en cientos de conversaciones exhaustivas y de alta calidad con representantes de fabricantes de baterías, fabricantes de equipos originales (OEM) de vehículos eléctricos, proveedores de componentes e instituciones de investigación de Norteamérica y de todo el mundo. Esta exposición no solo fue una exitosa presentación de la marca, sino también una valiosa experiencia de aprendizaje y conocimiento. Creemos firmemente que, con el rápido desarrollo de la industria de los vehículos eléctricos y la continua evolución de la tecnología de baterías, los requisitos de pureza y control de precisión en el entorno de producción son cada vez más estrictos. KLC aprovechará esta exposición como un nuevo punto de partida para aumentar continuamente la inversión en I+D y optimizar continuamente nuestros productos y tecnologías, con el objetivo de ofrecer soluciones de filtración de aire y salas blancas más seguras, eficientes y económicas para la cadena global de la industria de las nuevas energías. Esperamos transformar las ideas generadas durante la exposición en fructíferas colaboraciones futuras y trabajar con colegas de la industria para aportar el "poder puro de KLC" a impulsar el futuro de la movilidad ecológica.